What Researchers Did
The researchers measured blood flow and skin oxygen levels in the face and gums of ten healthy volunteers, repeating the tests over several weeks to see how consistent the readings were.
What They Found
This study of ten healthy adults found acceptable reproducibility for laser Doppler flowmetry (LDF) and transcutaneous oximetry (TcPO2), with week-to-week intraclass correlation coefficients of 0.74 to 0.96 for LDF and 0.44 to 0.75 for TcPO2. LDF readings taken after heat-induced vessel widening were more reproducible than baseline flow, in both facial skin and gum tissue. The researchers also saw pronounced variation within the same day. They recommended heated probes (thermoprobes) and repeating measurements at the same time of day.
Canadian Relevance
No direct Canadian connection identified.
Study Limitations
The study included only ten healthy volunteers, so the abstract's findings on reproducibility may not apply to patients with vascular disease or radiation-damaged tissue, and TcPO2 reproducibility was lower than LDF.